Optimization of a hematology laboratory: how smart settings save resources without losing diagnostic quality? 🔬
Modern hematology analyzers are high-tech systems that not only count cells but also signal pathological populations using so-called “flags” (alerts). However, standard factory settings are often too sensitive, leading to a huge number of false positives and the need to manually recheck results.
A group of researchers from Denmark conducted extensive work to optimize these parameters on the Sysmex XN-9100 platform, using flow cytometric differential (FlowDiff) as a reference method. 🧬
📊 What was done? During the study, data from 4554 blood samples were analyzed. Scientists revised the threshold values of Q-flags (quantitative parameters suspicious for pathology) and evaluated the clinical significance of counting flags.
Key changes in settings:
🔹 Q-flag optimization: Based on the Youden index, new, more accurate cut-off points were established:
- “Atypical lymphocytes” — threshold raised to 120 (instead of 100).
- “Abnormal lymphocytes” — to 170.
- “Blasts” (suspicion of immature cells) — to 160.
🔹 Elimination of redundancy: The “Left shift” flag (shift of the leukocyte differential to the left) showed no independent diagnostic value and was deactivated, as it duplicated other parameters.
🔹 Trust in automation: The flags “eosinophilia”, “basophilia”, and “monocytopenia” were also disabled. Comparison with reference flow cytometry confirmed that the analyzer provides highly accurate results for these items even when flags are present.
✅ Results and conclusions:
- Efficiency: The implementation of new rules allowed avoiding additional manual analysis in 37.5% of cases. This significantly reduced the workload on staff and shortened the Turnaround Time. ⏱
- Safety: Verification on a separate cohort of patients with acute leukemia confirmed: the new strategy has the same diagnostic sensitivity as the factory settings. The risk of missing a pathological process remained minimal. 🛡
- Savings: In the first 9 months of operation under the new rules alone, the laboratory saved over 64,000 euros by reducing unnecessary confirmatory tests. 💰
This study proves that individual customization of analytical systems for the tasks of a specific medical institution is not just a whim, but a way to make diagnostics faster and more economically sound, while maintaining the highest clinical accuracy. 🧪
Source for citation: Tougaard NH, Dimopoulos K, Pedersen SKV, Abildgaard A, Winther-Larsen A. Optimization of Leukocytic Flags in the Sysmex-XN Series With the Help of Flow Cytometric Differential (FlowDiff) Can Significantly Optimize Laboratory Efficiency. International Journal of Laboratory Hematology, 2026; 0:1–8. DOWNLOAD PDF Unlock the secrets of blood and bone marrow